SYNCHRON UNICEL DXC 600
Report
- Report Number
- 2050012-2011-00505
- Date Received
- March 1, 2011
- Date of Event
- January 28, 2011
- Report Date
- January 28, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- OTHER
Narratives
QC WAS WITHIN LABORATORY ESTABLISHED RANGES PRIOR TO THE EVENT. QC RUN IN THE AFTERNOON WAS OUT OF RANGE LOW. QC WAS RERUN AT 1734 WAS WITHIN RANGE. THE CUSTOMER REPLACED THE K TIP WHICH RESOLVED THE ISSUE. SERVICE WAS NOT DISPATCHED; HOWEVER, SERVICE ASSISTED AND INSTRUCTED THE CUSTOMER VIA TELEPHONE TO RUN ADDITIONAL SAMPLES TO STABILIZE THE NEW ELECTRODE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT AN ERRONEOUS POTASSIUM (K) RESULT GENERATED BY THE SYNCHRON UNICEL DXC 600 CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS RERUN AFTER TROUBLESHOOTING CORRECTED THE ISSUE AND HIGHER RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT INITIATED OR WITHHELD DUE TO THE LOW RESULT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON UNICEL DXC 600 | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC600 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |