FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 2003400 · Received March 1, 2011

Report

Report Number
9611820-2011-00008
Event Type
Injury
Date Received
March 1, 2011
Report Date
January 31, 2011
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ABUTMENT HEX FRACTURE WAS LIKELY DUE TO BIOMECHANICAL OVERLOADING DURING THE FIVE YEARS IN WHICH THE ABUTMENT WAS USED AS A POST FOR A TELESCOPIC RESTORATION IN THE REGION OF THE CANINE TEETH. AN EVALUATION OF THE RETURNED ABUTMENT INDICATED THAT THE ABUTMENT MET PRODUCT SPECIFICATIONS. IT MAY THEREFORE BE CONCLUDED THAT THE FRACTURE WAS NOT DUE TO A PRODUCT FAILURE. (B)(4).

Description of Event or Problem · 1

ON (B)(6), 2011 A DOCTOR REPORTED TO (B)(4) THAT A PITT EASY ABUTMENT FRACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 61 YR Other| R