FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2003184 · Received February 22, 2011

Report

Report Number
1824206-2011-01025
Event Type
Malfunction
Date Received
February 22, 2011
Date of Event
February 8, 2011
Report Date
February 8, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND BED IN "DOWN" STORAGE AREA. WHEN TESTING BED FOUND CPR FUNCTION IS NOT WORKING, HEAD DOWN FUNCTION WORKS BUT THE HEAD GOES DOWN VERY SLOWLY. AFTER TROUBLESHOOTING, DETERMINED PROBLEM TO BE A FAULTY CPR VALVE. REPLACED CPR VALVE TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

BED FOUND IN "DOWN" STORAGE AREA. CAUSE OF PROBLEM IS UNK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900G006241

Patients

Seq Age Sex Outcome Treatment
1