FDA Adverse Event Injury Summary report: N

ATUNE REV RP TIB BASE SZ 7 CEM

MDR report key: 20031802 · Received August 22, 2024

Report

Report Number
1818910-2024-17803
Event Type
Injury
Date Received
August 22, 2024
Date of Event
July 3, 2020
Report Date
August 21, 2024
Manufacturer
DEPUY IRELAND - 9616671
Product Code
NJL
UDI-DI
10603295042914
PMA / PMN Number
P830055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). H6 COMPONENT CODE: APPROPRIATE TERM/CODE NOT AVAILABLE (G07002) USED TO CAPTURE NO FINDINGS AVAILABLE. INVESTIGATION SUMMARY ==> NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A RECORDS EVALUATION (MRE) WAS NOT PERFORM AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT ==> THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

SUBJECT ID: (B)(6). STUDY: DSJ-2016-02 CLINICAL ADVERSE EVENT RECEIVED FOR INFECTION LEFT KNEE EVENT IS SERIOUS AND IS CONSIDERED SEVERE EVENT IS DEFINITELY NOT RELATED TO DEVICE AND IS DEFINITELY RELATED TO PROCEDURE. DATE OF IMPLANTATION: (B)(6) 2019 DATE OF EVENT (ONSET): 07/03/2020 (LEFT KNEE). TREATMENT: SURGICAL IRRIGATION & DEBRIDEMENT, WITH REVISION OF TIBIAL INSERT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23659 ATUNE REV RP TIB BASE SZ 7 CEM KNEE TIBIAL TRAY NJL DEPUY IRELAND - 9616671 8873302 10603295042914

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Required Intervention