FDA Adverse Event Injury Summary report: N

TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 2003113 · Received February 28, 2011

Report

Report Number
2134265-2011-00645
Event Type
Injury
Date Received
February 28, 2011
Date of Event
April 12, 2010
Report Date
February 1, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

(B)(4). SAME CASE AS MFR REPORT #: 2134265-2010-04589. IT WAS REPORTED THAT FOLLOWING A STENTING TREATMENT PROCEDURE, THE PATIENT PRESENTED WITH IN STENT RESTENOSIS. THE INDEX PROCEDURE TREATED THE 2.5X32MM TARGET LESION LOCATED IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. TREATMENT CONSISTED OF PRE-DILATION AND THE PLACEMENT OF TWO OVERLAPPING 2.5X32MM TAXUS LIBERTE STENTS RESULTING IN 0% RESIDUAL STENOSIS. A NON-TARGET LESION LOCATED IN THE PROXIMAL LAD WAS TREATED WITH A NON BSC STENT. AFTER SCHEDULED TREATMENT FOR ANOTHER LESION, THE PATIENT WAS DISCHARGED WITHOUT COMPLICATION 12 DAYS LATER ON ASPIRIN AND TICLOPIDINE. IN (B)(6) 2010, WITHOUT ISCHEMIC SYMPTOM QCA EVALUATION REVEALED A 79% RESTENOSED, 2.83X13.3MM TARGET LESION LOCATED IN THE MID LAD WITH TIMI 3 FLOW. TREATMENT UTILIZED A CUTTING BALLOON AND ANGIOPLASTY RESULTING IN 23% RESIDUAL STENOSIS AND TIMI 3 FLOW. THE PATIENT WAS DISCHARGED TWO DAYS POST THE PROCEDURE. NO ANGINA WAS CONFIRMED AT THE 9 MONTH AND 1 YEAR STUDY FOLLOW UP VISITS. PER THE PHYSICIAN, THE EVENT WAS LISTED AS "POSSIBLY RELATED" TO THE STUDY STENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493893632250

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention 2.5X32MM TAXUS LIBERTE| CYPHER