FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2003086 · Received February 8, 2011

Report

Report Number
2531779-2011-00694
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
January 8, 2011
Report Date
January 8, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE PUMP FELT HOT TO THE TOUCH. SHE STATED THAT SHE FELT THE PUMP DURING THE NIGHT GET VERY HOT TO TOUCH. THE PATIENT NOTED THAT THE SIDE OF THE PUMP NEAR THE BATTERY COMPARTMENT WAS VERY HOT, THE PUMP EMITTED A LOW BATTERY ALARM WHICH WAS CONFIRMED IN THE ALARM HISTORY. THE PATIENT CONFIRMED THAT THE BATTERY SHE WAS USING AT THE TIME OF THE INCIDENT WAS THE ENERGIZER LITHIUM BATTERY THAT WAS SENT WITH THE PUMP KIT; SHE RECEIVED THE PUMP THREE DAYS AGO WHEN SHE BEGAN INSULIN PUMP THERAPY. SHE CONFIRMED THAT THERE WERE NO SIGNS OF MOISTURE OR CORROSION ON THE BATTERY BUT THE BATTERY WAS WARM TO TOUCH. THE PATIENT CHANGED THE BATTERY TO AN ALKALINE BATTERY AND CONTINUED TO USE THE PUMP WITHOUT FURTHER INCIDENTS. THE PATIENT DENIED BODILY INJURY OR BURNS AS A CONSEQUENCE OF THE HEAT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 65 YR