FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2003071 · Received February 8, 2011

Report

Report Number
3004209178-2011-80353
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
January 12, 2011
Report Date
January 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED EXPERIENCING HIGH BLOOD GLUCOSE FOR THE PAST TWO DAYS. THE CUSTOMER CHANGED THE INFUSION SET AND RESERVOIR TO SOLVE THE PROBLEM. TROUBLESHOOTING WAS PERFORMED. THE SETTINGS, TIME, AND DATE ON THE INSULIN PUMP WERE CORRECT. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. PERFORMED A HIGH PRESSURE TEST AND THE DEVICE FAILED THE TEST. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 25 YR