FDA Adverse Event Malfunction Summary report: N

MAVERICK2 BALLOON CATHETER

MDR report key: 2003056 · Received February 28, 2011

Report

Report Number
2134265-2011-00375
Event Type
Malfunction
Date Received
February 28, 2011
Date of Event
February 1, 2011
Report Date
February 1, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE A BALLOON RUPTURE OCCURRED. ACCESS WAS GAINED THROUGH THE RADIAL ARTERY. THE LESION WAS LOCATED IN THE MID LEFT ANTERIOR DESCENDING ARTERY. THE 1.5X15MM MAVERICK 2 MONORAIL BALLOON CATHETER WAS ADVANCED TO THE TARGET LESION WHERE THE BALLOON RUPTURED AT 4 ATMS ON THE SECOND INFLATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAVERICK2 BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493892815150 0013722688

Patients

Seq Age Sex Outcome Treatment
1