FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2003036
·
Received February 8, 2011
Report
- Report Number
- 1824206-2011-00731
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- January 19, 2011
- Report Date
- January 19, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
HEAD IS SLOWLY DRIFTING DOWN, ABOUT 1 INCH PER 30 MIN. NO INCIDENT WAS REPORTED AS A RESULT OF THIS ISSUE. HEAD DOWN VALVE AND CPR VALVE WERE CHECKED. REPLACED THE CPR VALVE TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
COMPLAINT ALLEGED THAT THE HEAD OF THE BED IS SLOWLY DRIFTING DOWN, ABOUT 1 INCH PER 30 MIN. NO INCIDENT WAS REPORTED AS A RESULT OF THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900D004931 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |