FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2003036 · Received February 8, 2011

Report

Report Number
1824206-2011-00731
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
January 19, 2011
Report Date
January 19, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HEAD IS SLOWLY DRIFTING DOWN, ABOUT 1 INCH PER 30 MIN. NO INCIDENT WAS REPORTED AS A RESULT OF THIS ISSUE. HEAD DOWN VALVE AND CPR VALVE WERE CHECKED. REPLACED THE CPR VALVE TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT THE HEAD OF THE BED IS SLOWLY DRIFTING DOWN, ABOUT 1 INCH PER 30 MIN. NO INCIDENT WAS REPORTED AS A RESULT OF THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900D004931

Patients

Seq Age Sex Outcome Treatment
1