HOMECHOICE
Report
- Report Number
- 1423500-2011-02529
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Date of Event
- September 18, 2010
- Report Date
- February 7, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE: INSUFFICIENT DRAIN, FALSE EMPTY DETECT AND USE ERROR, INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).
(B)(4). DEVICE EVALUATION EXPECTED, BUT NOT YET COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.
AN INSTANCE OF INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING A REVIEW OF THE RETURNED HOMECHOICE (HC) LOG. ON (B)(6) 2010, THE DRAIN VOLUME WAS 2090ML DURING CYCLE 1. NO PATIENT INJURY OR MEDICAL INTERVENTION HAS BEEN REPORTED IN ASSOCIATION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |