FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2002992
·
Received February 8, 2011
Report
- Report Number
- 1824206-2011-00724
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- January 14, 2011
- Report Date
- January 14, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT DECLINED TO REPAIR THE BED AT THIS TIME.
Description of Event or Problem · 1
THE ACCOUNT FOUND THE TRENDELENBURG FUNCTION DOES NOT WORK ON THE BED. HE FOUND THE TREND LINKAGE BOLD WAS NOT PROPERLY ADJUSTED AND THE TREND/CPR VALVE RUBBER SEAT WAS DAMAGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |