FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2002982 · Received February 28, 2011

Report

Report Number
1423500-2011-02526
Event Type
Malfunction
Date Received
February 28, 2011
Date of Event
December 17, 2010
Report Date
February 7, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE PRODUCT ANALYSIS LAB EVALUATED THE DEVICE AND FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) DURING EVALUATION. BASED ON A REVIEW OF ALL AVAILABLE THERAPY LOG DATA, THE CAUSE OF THE IIPV WAS DETERMINED TO BE FALSE EMPTY DETECT / USE ERROR AS THE INITIAL DRAIN ALARM SETTING WAS INAPPROPRIATELY PROGRAMMED. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. LABELING REVIEW FOUND LABELING TO BE ADEQUATE FOR THE USER ERROR IDENTIFIED IN THIS REPORT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED BY BAXTER FOR EVALUATION. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 4. THE PATIENT'S DRAIN VOLUME WAS 4221ML. THE FILL VOLUME WAS 2500ML. THIS MEETS IIPV CRITERIA. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON 02/28/2011. ACCORDING TO THE NURSE THE PATIENT WAS SEEN LAST WEEK AND HAS RESUMED THERAPY SUCCESSFULLY. THE PATIENT DID NOT MENTION HAVING ANY ISSUES WITH THE CYCLER AND DID NOT REPORT ANY SYMPTOMS OF OVERFILL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 77 YR HOMECHOICE PRO