FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2002662 · Received February 9, 2011

Report

Report Number
2531779-2011-00710
Event Type
Malfunction
Date Received
February 9, 2011
Date of Event
January 10, 2011
Report Date
January 10, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. EVAL INDICATED THAT LOSS OF CARTRIDGE DETECTION COULD NOT BE VERIFIED IN THE PUMP HISTORY OR DUPLICATED DURING TESTING. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.

Description of Event or Problem · 1

THE PT REPORTED THAT ABOUT EVERY FIVE TIMES HE CHANGES HIS CARTRIDGE THE PUMP PUSHES ALL OF THE INSULIN OUT OF THE CARTRIDGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 35 YR