FDA Adverse Event Malfunction Summary report: N

ENDOSCOPIC APPLIER

MDR report key: 2002658 · Received February 28, 2011

Report

Report Number
3005075853-2011-00763
Event Type
Malfunction
Date Received
February 28, 2011
Date of Event
February 1, 2011
Report Date
February 2, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K830503
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). JAWS. THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED MAKING THE INSTRUMENT NON-FUNCTIONAL. A POSSIBLE CAUSE OF THIS CONDITION MAY BE THE INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS WHICH MAY RESULT IN CLIP MALFORMATION. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FOUND CONDITION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, WITH THIS APPLIER THE CLIP IS NOT GETTING CLOSED PROPERLY. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOSCOPIC APPLIER FZP ETHICON ENDO-SURGERY, LLC. NA AT7955

Patients

Seq Age Sex Outcome Treatment
1