FDA Adverse Event Malfunction Summary report: N

CESAR/C-STANDFLOOR-OMCP-VISUB-CCD (V5000)

MDR report key: 2002528 · Received January 28, 2011

Report

Report Number
3003768277-2011-00071
Event Type
Malfunction
Date Received
January 28, 2011
Report Date
December 29, 2010
Manufacturer
PHILIPS HEALTHCARE
Product Code
IZI
PMA / PMN Number
K984545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION FOUND THAT THERE WAS NO FIRE OR SMOKE, JUST A BURNING ODOR. THE FIELD SERVICE ENGINEER (FSE) FOUND THAT THERE WERE FOUR ISSUES WITH THE SYSTEM: THE POWER SUPPLY HAD A BROKEN CAPACITOR AND WAS REPLACED, ONE OF THE FOUR FANS RESPONSIBLE FOR COOLING THAT AREA OF THE CABINET WAS NOT WORKING AND WAS REPLACED. THE FSE INDICATED THAT, THAT WAS THE AREA OF THE CABINET WHERE THE ODOR WAS COMING FROM. THERE WAS A FAN IN THE POWER SUPPLY NEW VISUB THAT WAS NOT WORKING AND WAS REPLACED. WHEN THE FSE SWITCHED ON THE POWER SUPPLY VISUB, THE RELAY DID NOT START AND WAS REPLACED. AFTER THE PARTS WERE REPLACED IN THE SYSTEM, THE REPORTED ISSUES WERE RESOLVED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A STRONG BURNING ODOR COMING FROM THE SISCO CABINET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CESAR/C-STANDFLOOR-OMCP-VISUB-CCD (V5000) IZI PHILIPS HEALTHCARE 72248

Patients

Seq Age Sex Outcome Treatment
1