FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 20024157 · Received August 22, 2024

Report

Report Number
1213809-2024-00559
Event Type
Malfunction
Date Received
August 22, 2024
Date of Event
July 26, 2024
Report Date
September 4, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP MDR FOR DEVICE EVALUATION/CORRECTION. FOLLOWING SUBMISSION OF INITIAL MDR, THE CUSTOMER INFORMATION WAS NOT CORRECTLY REPORTED. SECTION E UPDATED TO REFLECT CORRECT CUSTOMER ADDRESS. EVALUATION: TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. BOTH PHOTOS SHOW WHITE PARTICULATES EMBEDDED IN THE BARREL. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE IMPROPERLY MOLDED BARREL IS ASSOCIATED WITH THE MOLDING PROCESS. THIS DEFECT IS OCCURRING BELOW AN EXPECTED RATE SO NO CORRECTIVE ACTIONS WILL BE MADE AT THIS TIME. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1082602. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHITE SPOTS WERE IDENTIFIED INSIDE A 1ML LUER LOCK SYRINGE WHILE RECONSTITUTION WAS BEING PERFORMED. IT LOOKED LIKE IT WAS EMBEDDED IN THE SYRINGE. THE SYRINGE HAS BEEN DISCARDED AS IT HAS BEEN USED TO RECONSTITUTE CYTOTOXIC DRUGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1397474 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 1082602 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown