DAVINCI XI

MDR report key: 20023798 · Received August 22, 2024

Report

Report Number
2955842-2024-18171
Event Type
Malfunction
Date Received
August 22, 2024
Date of Event
July 23, 2024
Report Date
July 23, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS VERIFIED AND READY TO USE. INTUITIVE SURGICAL, INC. (ISI) HAS ISSUED A RETURN MATERIAL AUTHORIZATION (RMA) TO HAVE THE IESU RETURNED. HOWEVER, ISI HAS NOT RECEIVED THE PRODUCT INVOLVED WITH THE ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE GENERATOR ASSOCIATED WITH THIS COMPLAINT AND COMPLETED INVESTIGATIONS. THE UNIT WAS RETURNED FOR FAILURE ANALYSIS AND THE REPORTED FAILURE WAS CONFIRMED AND REPRODUCED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INTEGRATED ELECTROSURGICAL UNIT (IESU) HAD ERROR C-00. AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER REVIEWED AND CONFIRMED SEVERAL ERRORS C-33 IN THE ERROR LOGS. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER, HOWEVER, NO FURTHER DETAILS COULD BE PROVIDED.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.