GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE
Report
- Report Number
- 2017233-2011-00071
- Event Type
- Injury
- Date Received
- February 16, 2011
- Date of Event
- January 14, 2011
- Report Date
- February 16, 2011
- Manufacturer
- W.L. GORE & ASSOCIATES, INC.
- Product Code
- DYB
- PMA / PMN Number
- K032073
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF THE MANUFACTURING PAPERWORK HAS BEEN CONDUCTED. THE REVIEW OF THE MANUFACTURING PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECIFICATIONS. ACCORDING TO THE GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE INSTRUCTIONS FOR USE, ENSURE THE VESSEL IS OF ADEQUATE SIZE AND APPROPRIATE TORTUOSITY FOR THE INSERTION OF THE INTRODUCER SHEATH. IF THE VESSEL IS TOO SMALL, MAJOR BLEEDING, VESSEL RUPTURE/PERFORATION, AND SERIOUS INJURY TO THE PATIENT INCLUDING DEATH MAY RESULT. DO NOT ATTEMPT TO ADVANCE OR WITHDRAW GUIDEWIRE, CATHETER, OR OTHER DEVICE THROUGH THE INTRODUCER SHEATH AND/OR DILATOR IF RESISTANCE IS FELT. USE FLUOROSCOPY TO DETERMINE THE CAUSE. CONTINUED ADVANCEMENT OR RETRACTION AGAINST RESISTANCE MAY RESULT IN SERIOUS INJURY TO THE PATIENT, DAMAGE TO OR BREAKAGE OF THE GUIDEWIRE, CATHETER, OR OTHER DEVICE.
ON (B)(6) 2011, THE PATIENT UNDERWENT REPAIR OF A THORACIC AORTIC ANEURYSM. A 24 FR GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE WAS ADVANCED WITH STRONG RESISTANCE. TWO GORE TAG THORACIC ENDOPROSTHESES WERE IMPLANTED. UPON REMOVAL OF THE 24 FR GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE, THERE WAS AGAIN STRONG RESISTANCE. THE PATIENT'S RIGHT EXTERNAL ILIAC ARTERY AND RIGHT INTERNAL ILIAC ARTERY WERE RUPTURED AND REPAIRED USING A DACRON GRAFT. THE PATIENT TOLERATED THE PROCEDURE. THE DIAMETER OF THE EXTERNAL ILIAC ARTERY WAS 7.9-8.2 MM. THE 24 FR GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE IS FOR A VESSEL DIAMETER OF 9.1 MM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE | DYB / INTRODUCER, CATHETER | DYB | W.L. GORE & ASSOCIATES, INC. | WLG | 8134787 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| R |