FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM

MDR report key: 2002365 · Received January 28, 2011

Report

Report Number
2024601-2011-00029
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
August 12, 2010
Report Date
January 5, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. MEDWATCH SENT TO FDA ON: (B)(4) 2011. VISUAL EXAM OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS SHOWED A THINNER SURFACE ON THE PORT TUBING AND A SMOOTH LINEAR OPENING WITH LEAKAGE CONSISTENT WITH WEAR AND TEAR. THE BAND TUBING WAS BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED A "PORT REMOVAL DUE TO A BIG LEAK IN THE PORT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM LTI ALLERGAN NA 1104585

Patients

Seq Age Sex Outcome Treatment
1 47 YR