FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 24GAX0.75IN PRN SLM NPVC

MDR report key: 20023245 · Received August 21, 2024

Report

Report Number
3002601200-2024-00391
Event Type
Malfunction
Date Received
August 21, 2024
Date of Event
July 29, 2024
Report Date
September 11, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830787
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE CUSTOMER RETURNED 1 PHOTO, NO DEFECTIVE SAMPLE. THE PHOTO SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. DHR/BHR REVIEW LOT#4052015. THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN APRIL 2024, AND PACKAGED AT R240 PACKAGE LINE IN APRIL 2024. WORK ORDER QUANTITY WAS (B)(4) EA. REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MET THE PRODUCT SPECIFICATIONS. REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. IN RESPONSE TO THE LEAKAGE AT THE SEPTUM, THE PLANT HAS LAUNCHED AN IN DEPTH INVESTIGATION THE 800MM SIMULATED CLINICAL LEAKAGE TEST WAS CONDUCTED ON THE RETAINED SAMPLES FROM THIS BATCH. IT WAS FOUND THAT A FEW SAMPLES LEAKED AT THE END OF THE SEPTUM AFTER THE NEEDLE TIP WAS INSERTED INTO THE TEST DEVICE, AND THE LEAKAGE STOPPED AFTER THE NEEDLE CORE WAS PULLED OUT. CONCLUSION(S): THE RETURNED PHOTO SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. IN RESPONSE TO THIS DEFECT, THE PLANT HAS LAUNCHED AN IN DEPTH INVESTIGATION TO TRACE AND INVESTIGATE ITS ROOT CAUSE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II 24GAX0.75IN PRN SLM NPVC LEAKED. THIS PRODUCT IS LEAKING BLOOD FROM THE END DURING USE; AFFECTED QUANTITY 1 UNIT, SAMPLE NOT RETURNABLE, PHOTOS AVAILABLE; GREEN CLAIM REQUIRED, COMPLAINT RESPONSE LETTER REQUIRED, COMPLAINT RECEIPT LETTER NOT REQUIRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1127416 BD INTIMA-II 24GAX0.75IN PRN SLM NPVC INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 4052015 00382903830787

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown