OCTRODE LEAD KIT, 60CM LENGTH

MDR report key: 2002323 · Received January 27, 2011

Report

Report Number
1627487-2011-00121
Event Type
Malfunction
Date Received
January 27, 2011
Date of Event
December 30, 2011
Report Date
December 30, 2011
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION DIVISION
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DEVICE HISTORY RECORD FOUND A NONCONFORMANCE; HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PT REC'D HER SCS SYSTEM ON (B)(6) 2010 INCLUDING TWO PERCUTANEOUS LEADS. IMPEDANCE MEASUREMENTS ON THE DEVICES WERE TAKEN INTRAOPERATIVELY, AND IT WAS REPORTED THAT ONE OF THE 16 CONTACTS YIELDED AN INVALID READING. NO ACTION WAS TAKEN DURING THE SURGICAL PROCEDURE TO CORRECT THIS ISSUE. POSTOPERATIVE, THE ALLEGED IMPEDANCE ISSUE REMAINED; HOWEVER, STIMULATION WAS CAPTURED FOR THE PT. THIS REPORT IS BEING SUBMITTED AS A PRECAUTIONARY MEASURE AS SIMILARLY REPORTED EVENTS HAVE LED TO THE EXPLANT OF THE LEAD (S).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OCTRODE LEAD KIT, 60CM LENGTH SPINAL CORD STIMULATION LEAD LGW ST JUDE MEDICAL - NEUROMODULATION DIVISION 3186 3240575

Patients

Seq Age Sex Outcome Treatment
1