I-STAT CG4+ CARTRIDGE
Report
- Report Number
- 2245578-2024-00165
- Event Type
- Malfunction
- Date Received
- August 21, 2024
- Date of Event
- August 15, 2024
- Report Date
- October 1, 2024
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- KHP
- PMA / PMN Number
- K200492
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 27-SEP-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CG4+ CARTRIDGE LOT M24076.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 15-AUG-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG4+ CARTRIDGE THAT YIELDED A SUSPECTED DISCREPANT LACTATE RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD, DATE, COLLECTED, TESTED, PH, PCO2, PO2, SO2, LACTATE, SAMPLE: I-STAT, (B)(6) 2024, NI, 06:08, 7.29, 51.90 MMHG, 15 MMHG, 16.0%, 0.7 MMOL/L, A; OPTI CCA-TS2, (B)(6) 2024, 07:00, 07:20, 7.32, 38 MMHG, 86 MMHG, 96%, 1.71MMOL/L, NI. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1300352 | I-STAT CG4+ CARTRIDGE | CG4+ CARTRIDGE | KHP | ABBOTT POINT OF CARE | NA | M24076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Female |