CE MULTIRATE INFUSOR LV2,3,512 PACK
Report
- Report Number
- 6000001-2011-01390
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Date of Event
- December 11, 2010
- Report Date
- December 13, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K011317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GR
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL EXAMINATION WAS PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A BROKEN DEVICE. THE ROOT CAUSE WAS DETERMINED TO BE A RUPTURED RESERVOIR. THE DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH IDC-CAPA-(B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF A MULTI-RATE INFUSOR DEVICE BROKE DURING FILLING. THE DEVICE WAS BEING FILLED WITH ANALGESIC SOLUTION WHEN THE DEVICE BROKE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE MULTIRATE INFUSOR LV2,3,512 PACK | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10G028 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ANALGESIC SOLUTION |