FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8 ML

MDR report key: 2002165 · Received February 4, 2011

Report

Report Number
3004209178-2011-80325
Event Type
Malfunction
Date Received
February 4, 2011
Date of Event
January 20, 2011
Report Date
January 20, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND. TROUBLESHOOTING WAS PERFORMED. RAN A DISPLACEMENT TEST AND THE DEVICE DID NOT PASS THE TEST. IT WAS STATED THAT AFTER THE RESERVOIR WAS CHANGED THE INSULIN PUMP FUNCTIONS PROPERLY. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESERVOIR 1.8 ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-326A H7744208

Patients

Seq Age Sex Outcome Treatment
1