PHASEAL INFUSION ADAPTER
Report
- Report Number
- 9615998-2011-00001
- Event Type
- Malfunction
- Date Received
- February 1, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 31, 2011
- Manufacturer
- CARMEL PHARMA AB
- Product Code
- LHI
- PMA / PMN Number
- K023747
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED INCIDENT DOES NOT INVOLVE DEATH OR SERIOUS INJURY TO PATIENT, USER OR OTHER PERSONS. THE SPIKE OF THE INFUSION ADAPTER IS MADE OF ABS PLASTIC (STATED IN INSTRUCTION FOR USE). SOME DRUGS OR DRUG SOLVENTS ARE KNOWN TO COMPROMISE THE INTEGRITY OF ABS PLASTIC. INVESTIGATIONS PERFORMED IN RELATION TO PREVIOUS REPORTS HAVE SHOWN THAT FRACTURES OF THE PHASEAL INFUSION ADAPTER C100 SPIKE MAY OCCUR AS THE RESULTS OF INTERACTION BETWEEN THE INFUSION ADAPTER SPIKE, CERTAIN DRUGS AND CERTAIN IV CONTAINER PORTS. A TECHNICAL BULLETIN WITH THIS INFORMATION, TOGETHER WITH A RECOMMENDATION TO FLUSH THE INFUSION ADAPTER WITH DILUTED DRUG IMMEDIATELY AFTER INJECTION HAS BEEN ISSUED TO ALL US CUSTOMERS AS A RESULT OF EARLIER INVESTIGATIONS. THE INSTRUCTION FOR USE HAS BEEN REVISED ACCORDINGLY.
INCIDENT REPORTED REGARDING INFUSION ADAPTER C100 FRACTURE WHEN PREPARED IN COMBINATION WITH BAG CONTAINING TAXOL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PHASEAL INFUSION ADAPTER | INTRAVASCULAR ADMINISTRATION SET | LHI | CARMEL PHARMA AB | C100 | 10060785 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |