FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2002088 · Received February 4, 2011

Report

Report Number
1824206-2011-00653
Event Type
Malfunction
Date Received
February 4, 2011
Date of Event
January 23, 2011
Report Date
January 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND LOOSE PINS IN P379 CABLE CONNECTOR. REPLACED P379 CABLE (PN (B)(4)) AND CHECKED FUNCTIONS TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT PT IN MEDSURG ROOM CAN PLACE NURSE CALL WITH RESPONSE NOT HEARD IN EXPECTED LOCATION. NO INJURY ALLEGED BY ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900B000153

Patients

Seq Age Sex Outcome Treatment
1