FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2002084 · Received February 4, 2011

Report

Report Number
1824206-2011-00657
Event Type
Malfunction
Date Received
February 4, 2011
Date of Event
January 12, 2011
Report Date
January 12, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE HEAD HI/LOW UP VALVE WAS NOT FUNCTIONING. HE REPLACED THE VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD HI/LOW FUNCTION IS DRIFTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1