HOMECARE HUMIDIFIER
Report
- Report Number
- 9611451-2011-00102
- Event Type
- Malfunction
- Date Received
- February 27, 2011
- Report Date
- January 31, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K953392
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT HC500JHU HOMECARE HUMIDIFIER AND THE ASSOCIATED BREATHING CIRCUITS HAVE NOT BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INSPECTION. OUR ANALYSIS IS BASED ON THE INFORMATION PROVIDED BY THE HOSPITAL AND THE PATIENT'S MOTHER. RESULTS: INFORMATION RECEIVED FROM THE HOSPITAL AND THE PATIENT'S MOTHER REVEALED THAT THE REPORTED BREATHING CIRCUITS USED WITH THE HC500JHU HUMIDIFIER THAT MELTED WERE MANUFACTURED BY CAREFUSION. IT WAS FURTHER REPORTED THAT "EXCESS CONDENSATION" WAS OBSERVED WHILE THE BREATHING CIRCUITS WERE BEING USED ON THE PATIENT. THE MOTHER OF THE PATIENT ALSO CONFIRMED, THROUGH THE LETTER THAT SHE SENT TO THE HOSPITAL, THAT THEY ALWAYS HAVE THE "TOP 2/3 OF THE CIRCUIT UNDER THE BEDDING. HOWEVER, THE CIRCUIT INSERT "CLEARLY STATES NOT TO COVER ANY PORTION OF THE CIRCUIT." CONCLUSION: CONDENSATE IN THE HUMIDIFICATION SYSTEM, ALTHOUGH NOT PREFERRED, IS AN EXPECTED SIDE EFFECT OF HEATED PASS-OVER HUMIDIFICATION SYSTEMS IN MANY CONDITIONS, AND MAY VARY BETWEEN LIGHT MISTING TO WATER DROPLETS THAT FORM ON THE WALL OF COOL BREATHING CIRCUIT TUBING. THERE ARE MANY FACTORS WHICH CAN RESULT IN CONDENSATION BUILD UP. THESE INCLUDE EQUIPMENT SET UP ERRORS, SUCH AS INCORRECT POSITIONING OF THE BREATHING CIRCUIT, OR EFFECTS OF ENVIRONMENTAL FACTORS (E.G. AIR CONDITIONING). IT IS ADVISED THAT WHEN SETTING UP HUMIDIFIED BREATHING CIRCUITS ENSURE THAT THE INSTRUCTIONS FOR USE ARE FOLLOWED AND THAT THE LOCAL ENVIRONMENT IS CONSIDERED. THE FISHER & PAYKEL HC500 RESPIRATORY HUMIDIFIER IS DESIGNED FOR USE WITH ARTIFICIAL VENTILATION SYSTEMS. THE DEVICE USER INSTRUCTIONS PROVIDE THE FOLLOWING WARNINGS: USE ONLY FISHER & PAYKEL APPROVED CHAMBERS, CIRCUITS AND ACCESSORIES. PERFORMANCE AND SAFETY CANNOT BE GUARANTEED IF OTHER TYPES OF ACCESSORIES ARE USED. DO NOT COVER THE PATIENT DELIVERY TUBE WITH BLANKETS, TOWELS OR SIMILAR MATERIAL AS THIS MAY CAUSE THE PATIENT DELIVERY TUBE TO OVERHEAT. TO PREVENT THE POSSIBILITY OF ACCIDENTAL PATIENT BURNS, ENSURE THAT THE HEATED BREATHING CIRCUIT IS NOT IN CONTACT WITH THE PATIENT'S SKIN.
(B)(4). THE COMPLAINT (B)(4) HOMECARE HUMIDIFIER IS NOT EXPECTED TO BE RETURNED TO FISHER & PAYKEL FOR INVESTIGATION. WE ARE CURRENTLY IN THE PROCESS OF OBTAINING FURTHER INFORMATION IN ORDER TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE EVENT AS REPORTED. WE WILL PROVIDE A FOLLOW UP REPORT ONCE WE RECEIVE FURTHER INFORMATION AND HAVE COMPLETED OUR ANALYSIS.
A PATIENT'S MOTHER REPORTED TO A RESPIRATORY THERAPIST IN PEYTON MANNING CHILDREN'S HOSPITAL THAT THEIR (B)(4) HOMECARE HUMIDIFIER "CAUSED SEVERAL BREATHING CIRCUITS TO BE MELTED". IT WAS FURTHER REPORTED THAT EXCESS CONDENSATION OCCURRED IN THE BREATHING CIRCUITS DURING USE.
A PATIENT'S MOTHER REPORTED TO A RESPIRATORY THERAPIST IN (B)(6) THAT THEIR HC500JHU HOMECARE HUMIDIFIER "CAUSED SEVERAL BREATHING CIRCUITS TO BE MELTED". IT WAS FURTHER REPORTED THAT EXCESS CONDENSATION OCCURRED IN THE BREATHING CIRCUITS DURING USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECARE HUMIDIFIER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | HC500JHU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |