FDA Adverse Event Other Summary report: N

DA VINCI SI SURGICAL SYSTEM

MDR report key: 2001958 · Received February 27, 2011

Report

Report Number
2955842-2011-00058
Event Type
Other
Date Received
February 27, 2011
Date of Event
January 26, 2011
Report Date
January 26, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081137
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THAT THE VISION ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE CAMERA CABLE. THE CAMERA CABLE CONNECTS THE CAMERA TO THE SYSTEM'S VISION CART, WHICH THEN TRANSMITS THE IMAGE TO THE SURGEON'S CONSOLE. THE SYSTEM WAS REPAIRED BY REPLACING THE DEFECTIVE CAMERA CABLE. THE DA VINCI SI SURGICAL SYSTEM USER'S MANUAL EXPLICITLY STATES THAT, ENVIRONMENTAL OR EQUIPMENT FAILURES MAY CAUSE THE DA VINCI SI SYSTEM TO BECOME UNAVAILABLE. THE SURGICAL TEAM SHOULD ALWAYS HAVE BACKUP EQUIPMENT AND INSTRUMENTATION AVAILABLE, AND BE PREPARED TO CONVERT TO ALTERNATIVE SURGICAL TECHNIQUES. AS OF FEBRUARY 24, 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI MITRAL VALVE REPLACEMENT PROCEDURE, THE SITE EXPERIENCED A BLACK IMAGE IN THE LEFT EYE OF THE SURGEON SIDE CONSOLE. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER, THE SURGICAL STAFF RESEATED THE CAMERA CABLE; HOWEVER, THE VISION ISSUE WAS NOT RESOLVED. AT THIS TIME, THE SURGICAL STAFF MADE THE DECISION TO CONVERT THE PLANNED PROCEDURE TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT HARM OR ADVERSE OUTCOME WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS3000 A6.0P4

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SI SYS, INSTRUMENTS AND ACCESSORIES