PRECISION XTRA
Report
- Report Number
- 2954323-2011-01754
- Event Type
- Injury
- Date Received
- February 27, 2011
- Date of Event
- December 24, 2010
- Report Date
- April 22, 2011
- Product Code
- NBW
- Removal / Correction Number
- ADC 1197-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
PRODUCT TESTING ON RETURNED STRIP LOT (45001A521) AND RETAINED METER ((B)(4)) DID NOT CONFIRM. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE RESULTS ARE CONSISTENT WITH THE RETAIN TESTING CONDUCTED, WHICH FOUND THAT NOT ALL STRIPS IN AFFECTED VIALS ARE IMPACTED. THIS IS A FINAL REPORT.
THIS IS AN INITIAL REPORT. THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30 DEGREES CELSIUS DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE "C", "D" OR "E" ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. THE FDA HAS BEEN INFORMED OF THE FIELD ACTION PER 21CFR806 (FIELD ACTION REFERENCE NUMBER 2954323-12/22/10-001-R). ALL AFFECTED CONSIGNEES WERE NOTIFIED BY LETTER BEGINNING DECEMBER 22, 2010.
IN USING AN AFFECTED TEST STRIP RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT'S PRECISION FAMILY TEST STRIPS, THE CUSTOMER REPORTED EITHER LONGER BLOOD FILL TIME OR READINGS ISSUES. A CUSTOMER REPORTED GETTING A LOWER THAN FEELS READING OF 67 MG/DL, FEELING JITTERY AND SLUGGISH, LOSING CONSCIOUSNESS AND REPORTEDLY "TOOK NOTHING SPECIAL" TO COUNTERACT THE EVENT. NO THIRD-PARTY MEDICAL INTERVENTION WAS REQUIRED. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS MEDICAL EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 45001A521 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |