FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 2001877 · Received February 25, 2011

Report

Report Number
2122870-2011-00500
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
January 27, 2011
Report Date
January 27, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K922823/A007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS LITHIUM HEPARIN PLASMA THAT WAS ASPIRED FROM THE PRIMARY COLLECTION TUBE FOR THE INITIAL RESULT. FOR THE REPEAT TESTING THE SAMPLE WAS ASPIRED FROM AN INSERT CUP. NO SAMPLE QUALITY ISSUE WAS NOTED. QC WAS WITHIN THE ESTABLISHED RANGES PRIOR TO THE EVENT. A SYSTEM CHECK WAS PERFORMED PRIOR TO THE EVENT AND THE RESULTS BARELY MET THE SPECIFICATIONS OF THE WASHED %CV, AT 11.18%. THE CUSTOMER PUT ON A DIFFERENT CLEAN SET OF ASPIRATE PROBES AND REPEATED THE SYSTEM CHECK. ALL RESULTS MET THE SPECIFICATIONS WITH THE WASHED %CV AT 5.82%. NO ERROR WAS POSTED TO THE EVENT LOG. SERVICE WAS ON SITE ON (B)(6) 2011 AND PERFORMED A PREVENTIVE MAINTENANCE. NO HARDWARE ISSUE WAS NOTED. NO DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS BEEN DETERMINED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) IN REGARDS TO AN ELEVATED TROPONIN (ACCUTNI) RESULT GENERATED BY ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT'S SAMPLE. REPEAT ANALYSIS ON THE SAME INSTRUMENT PRODUCED RESULTS WITHIN THE RISK STRATIFICATION RAGE. THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. SUBSEQUENT TESTING BY AN ALTERNATE METHOD PRODUCED A RESULT WITHIN THE NORMAL REFERENCE RANGE. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS® 2 IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1