FDA Adverse Event Injury Summary report: N

PROSTAR XL SUTURE-MEDIATED CLOSURE

MDR report key: 2001733 · Received February 25, 2011

Report

Report Number
2024168-2011-01207
Event Type
Injury
Date Received
February 25, 2011
Date of Event
January 28, 2011
Report Date
February 1, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (CONCOMITANT MEDICAL PRODUCTS): SHEATH: 24 FR OTHER: TAG THORACIC ENDOPROSTHESIS(GORE), INTRODUCER GORE 24 FR. (B)(4) - FAILURE TO FOLLOW STEPS/INSTRUCTIONS. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN, TRAINED IN THE USE OF THE PROSTAR XL DEVICE, ATTEMPTED ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY (CFA) AFTER PLACEMENT OF A THORACIC ENDOPROSTHESIS. REPORTEDLY, DURING PRE-CLOSURE, THE PROSTAR DEVICE WAS PLACED, THE NEEDLES DEPLOYED AND THE SUTURES WERE LEFT ASIDE. THE GRAFT WAS PLACED IN THE AORTA AND THE 24 FR SHEATH WAS REMOVED. VESSEL CLOSURE WAS ATTEMPTED BY ADVANCING THE KNOTS WITHOUT COMPLICATIONS, HOWEVER ARTERIAL BLEEDING WAS STILL OBSERVED. A DECISION WAS MADE TO SURGICALLY CLOSE THE ACCESS SITE, AND AT THAT TIME, A TEAR AT THE CFA WAS NOTICED. THE TEAR AND THE ACCESS SITE WERE REPAIRED/SUTURED USING A PROLENE SUTURE. THE VASCULAR SURGEON INDICATED THAT THE TEAR WAS ATTRIBUTED TO THE 24 FR SHEATH REMOVAL FROM THE ARTERIOTOMY AND RULED OUT THAT THE VESSEL CLOSURE DEVICE CONTRIBUTED TO IT. A FEMORAL ANGIOGRAM WAS NOT TAKEN. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROSTAR XL SUTURE-MEDIATED CLOSURE SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 900036H

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention