FDA Adverse Event Malfunction Summary report: N

RESTORE ULTRA

MDR report key: 2001544 · Received January 26, 2011

Report

Report Number
3004209178-2011-00636
Event Type
Malfunction
Date Received
January 26, 2011
Date of Event
January 1, 2011
Report Date
January 12, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

RECEIVED INFO, THE PT WAS UNABLE TO ADJUST STIMULATION, REPORTS SEEING THE EMPTY PT PROGRAMMER BATTERIES. NOT ABLE TO ADJUST STIMULATION WITH OR WITHOUT THE ANTENNA. ALSO NOT ABLE TO RECHARGE THE INS. MEDTRONIC REP REPORTS DEVICE IS SHOWING A POWER ON RESET MODE. BATTERY IS SHOWING EMPTY, BUT NOT OVERDISCHARGED. PT WAS INSTRUCTED TO CHARGE UNTIL 1/4 FULL AND THEN REP WILL CLEAR THE POR. ADD'L INFO HAS BEEN REQUESTED AND IF RECEIVED, A FOLLOW-UP REPORT WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ULTRA LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37712 NA

Patients

Seq Age Sex Outcome Treatment
1 45 YR EXTENSION: MODEL 37081, LOT# NJB055723V| EXPLANTED:| IMPLANTED:| PROGRMMMER: MODEL 37743, LOT# NKE129614N| ACCESSORY: MODEL 37752, LOT# NKA142958N| EXPLANTED:| IMPLANTED:| IMPLANTED:| EXTENSION: MODEL 37081, LOT# NJB053474V| EXPLANTED:| LEAD: MODEL 39286-30, LOT# N210570001