FDA Adverse Event Malfunction Summary report: N

HEARTMATE II SYSTEM CONTROLLER

MDR report key: 2001525 · Received January 27, 2011

Report

Report Number
2916596-2011-00041
Event Type
Malfunction
Date Received
January 27, 2011
Date of Event
October 25, 2010
Report Date
January 4, 2011
Manufacturer
THORATEC CORP.
Product Code
DSQ
Removal / Correction Number
29165969/2/10001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM CONTROLLER WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE REPORTED EVENT WAS CONFIRMED. DURING FUNCTIONAL TESTING OF THE RETURNED SYSTEM CONTROLLER, NO ALARM CONDITIONS WERE OBSERVED; HOWEVER, WHEN THE BLACK POWER LEAD WAS MANEUVERED AT THE CONNECTOR END DURING TESTING, THE SYSTEM REPORTED LOW VOLTAGE ALARMS AND A "REPLACE SYSTEM CONTROLLER ADVISORY" MESSAGE AND THEN THE PUMP STOPPED. IN ADDITION, WHEN THE SYSTEM WAS OPERATED SOLELY ON THE BLACK POWER LEAD, THE SYSTEM CONTROLLER REPORTED RED HEART AND YELLOW BATTERY ALARMS. THE SYSTEM CONTROLLER WAS DISCONNECTED FROM THE TEST SYSTEM AND UPON INSPECTION OF THE INNER CONDUCTORS AT THE CONNECTOR END OF THE BLACK POWER LEAD, THE BROWN AND BLUE WIRE CONDUCTORS WERE COMPROMISED. THIS SITUATION IS BEING ADDRESSED THROUGH THE MANUFACTURER'S CORRECTIVE/PREVENTIVE ACTION SYSTEM AND AN URGENT MEDICAL DEVICE CORRECTION NOTICE (29165969/2/10001-C) WAS SENT TO CUSTOMERS. NO FURTHER INFORMATION IS AVAILABLE. THE MANUFACTURER IS CLOSING ITS FILE ON THIS EVENT.

Description of Event or Problem · 1

THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). THE VAD COORDINATOR REPORTED THAT THE PATIENT WAS EXPERIENCING UNIDENTIFIED ALARMS. THE SUSPECT SYSTEM CONTROLLER WAS EXCHANGED PER THE MANUFACTURER'S INSTRUCTIONS FOR USE AND THE PATIENT REMAINS ONGOING ON LVAD SUPPORT WITH NO FURTHER ISSUE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II SYSTEM CONTROLLER LVAD SYSTEM CONTROLLER DSQ THORATEC CORP. 103696 NA

Patients

Seq Age Sex Outcome Treatment
1 31 YR Other