HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-02454
- Event Type
- Malfunction
- Date Received
- February 25, 2011
- Date of Event
- December 22, 2010
- Report Date
- February 3, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). BAXTER CONTACTED THE NURSE TO NOTIFY OF THIS INCIDENT OF INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS FOUND DURING THE DEVICE LOG REVIEW. THE HOMECHOICE (HC) DEVICE WAS RECEIVED FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE IIPV EVENT. DEVICE FUNCTIONED NORMALLY DURING PAL EVALUATION. THE CAUSE FOR THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO USE ERROR; THE TIDAL ULTRAFILTRATE REMOVAL WAS SET TOO LOW. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE SUFFICIENT. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THIS IS REPORT 1 OF 2 ASSOCIATED WITH THIS DEVICE. THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED ON (B)(6) 2010 DURING DRAIN CYCLE 5. THE PATIENT'S LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 1500 ML AND THE DRAIN VOLUME WAS 2428 ML, WHICH MET IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |