FDA Adverse Event Malfunction Summary report: N

ECHELON* 60 ENDOPATH

MDR report key: 2001451 · Received February 25, 2011

Report

Report Number
3005075853-2011-00747
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
January 28, 2011
Report Date
February 2, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). KNIFE. THE LONG60 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST CARTRIDGE RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE OPENED AND CLOSED WITHOUT DIFFICULTIES. THE CUT WAS NOTED TO BE JAGGED DUE TO A DAMAGED KNIFE. ONE POSSIBLE CAUSE FOR THE DAMAGED KNIFE MAY BE IF THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. IT IS ALSO RECOMMENDED THAT PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY PROCEDURE, THE STAPLER DID NOT OPEN AFTER THE FIRST FIRING AND HAS TO BE OPENED WITH THE HELP OF OTHER INSTRUMENTS. THE SURGEON HAS TO FINISH THE CASE WITH THE HELP OF ANOTHER STAPLER. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON* 60 ENDOPATH STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK G4U684

Patients

Seq Age Sex Outcome Treatment
1