FDA Adverse Event Malfunction Summary report: N

ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

MDR report key: 2001448 · Received February 25, 2011

Report

Report Number
1030489-2011-00204
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
January 31, 2011
Report Date
January 31, 2011
Manufacturer
MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO DETERMINE THE CAUSE OF THE EVENT. DEVICE HISTORY RECORDS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THE INFERIOR SHAFT IS BROKEN AT THE CENTERLINE OF THE STATIC JAW PIVOT PIN. OPTICAL EXAMINATION OF FRACTURE SURFACE DISCOVERED A FAIRLY BRITTLE FRACTURE, WITH NO INDICATION OF FATIGUE. THE LOCATION AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFTS ARE CONSISTENT WITH APPLICATION OF EXCESSIVE FORCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT WAS BROKEN BUT NOT BROKEN OFF DURING USE. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHOPEDIC MANUAL SURGICAL INSTRUMENT ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING NA GZ09G048

Patients

Seq Age Sex Outcome Treatment
1