OPTEASE VENA CAVA FILTER
Report
- Report Number
- 9610978-2011-00038
- Event Type
- Injury
- Date Received
- February 25, 2011
- Report Date
- January 31, 2011
- Manufacturer
- CORDIS EUROPA, N.V.
- Product Code
- DTK
- PMA / PMN Number
- K034050
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
COMPLAINT CONCLUSION: AN OPTEASE FILTER COULD NOT BE REMOVED DUE TO THROMBUS. THE FILTER WAS PLACED IN THE PATIENT AND THE PATIENT RETURNED FOR FILTER REMOVAL THREE WEEKS LATER. REMOVAL COULD NOT BE PERFORMED BECAUSE THE FILTER, INFERIOR VENA CAVA, AND ILIAC VEINS WERE OCCLUDED WITH THROMBUS. MULTIPLE ATTEMPTS HAVE BEEN MADE TO GATHER ADDITIONAL INFORMATION. HOWEVER, NO ADDITIONAL INFORMATION REGARDING PATIENT, LESION OR PROCEDURAL CHARACTERISTICS REGARDING THIS EVENT HAS BEEN PROVIDED. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. ADDITIONALLY, AS THE STERILE LOT NUMBER WAS NOT AVAILABLE, DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED. BASED ON THE LACK OF INFORMATION AND THE INABILITY TO ASSIGN OR DETERMINE A ROOT CAUSE NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THROMBOSIS IN THE FILTER IS A WELL KNOWN POTENTIAL COMPLICATION AND OCCURS IN APPROXIMATELY 3.6 TO 11.2 % OF ALL PATIENT' AND IS LISTED IN THE IFU AS SUCH. THE IFU ALSO STATES THAT RETRIEVAL OF THE OPTEASE FILTER SHOULD NOT BE ATTEMPTED IF THROMBUS IS PRESENT IN THE FILTER AND/OR CAUDAL TO THE FILTER. IT IS RECOGNIZED THAT THROMBI USUALLY DEVELOP FIRST IN THE CALF VEINS, "GROWING" IN THE DIRECTION OF FLOW OF THE VEIN. DVTS ARE DISTINGUISHED AS BEING ABOVE OR BELOW THE POPLITEAL VEIN. VERY EXTENSIVE DVTS CAN EXTEND INTO THE ILIAC VEINS OR THE INFERIOR VENA CAVA. DVT OCCURS WHEN A BLOOD CLOT FORMS IN A DEEP VEIN AND IS MOST COMMON IN THE DEEP VEINS OF YOUR LOWER LEG (CALF), AND CAN SPREAD UP TO THE VEINS IN YOUR THIGH. DVT CAN ALSO FIRST DEVELOP IN THE DEEP VEINS IN YOUR THIGH AND, MORE RARELY, IN OTHER DEEP VEINS, SUCH AS THE ONES IN YOUR ARM. DEEP VEIN THROMBOSIS IS THE RESULT OF THREE PRINCIPAL FACTORS : REDUCED OR STAGNANT BLOOD FLOW IN DEEP VEINS (VENOUS STASIS). INJURY TO THE BLOOD VESSEL WALL. AN INCREASE IN THE ACTIVITY OF THOSE SUBSTANCES IN THE BLOOD THAT ARE PART OF THE NORMAL CLOTTING MECHANISM, A CONDITION CALLED HYPERCOAGULABILITY (WHICH MEANS A MORE ACTIVE CLOTTING STATE). PLACEMENT OF A VENA CAVA FILTER IS NOT A CURE FOR DVT NOR DOES IT PREVENT THE FORMATION OF DVT. INFERIOR VENA CAVA FILTERS ARE USED TO PREVENT SEQUELAE, ESPECIALLY PE, IN PATIENTS WITH CONTRAINDICATIONS TO, COMPLICATIONS OF, OR FAILURE OF ANTICOAGULATION THERAPY AND PATIENTS WITH EXTENSIVE FREE-FLOATING THROMBI OR RESIDUAL THROMBI FOLLOWING MASSIVE PE. FACTORS THAT MAY HAVE INFLUENCED THE EVENT INCLUDE PATIENT, PHARMACOLOGICAL AND LESION.
THE LOT NUMBER COULD NOT BE PROVIDED. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
AN OPTEASE FILTER COULD NOT BE REMOVED DUE TO THROMBUS. THE FILTER WAS PLACED IN THE PATIENT AND THE PATIENT RETURNED FOR FILTER REMOVAL THREE WEEKS LATER. REMOVAL COULD NOT BE PERFORMED BECAUSE THE FILTER, INFERIOR VENA CAVA, AND ILIAC VEINS WERE OCCLUDED WITH THROMBUS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OPTEASE VENA CAVA FILTER | THROMBECTOMY SYSTEMS (DTK) | DTK | CORDIS EUROPA, N.V. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |