COBAS E411 DISK SYSTEM
Report
- Report Number
- 1823260-2011-01034
- Event Type
- Malfunction
- Date Received
- February 25, 2011
- Date of Event
- February 7, 2011
- Report Date
- May 13, 2011
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- MMI
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
BASED UPON THE INFORMATION PROVIDED, BOTH RESULTS WERE WITHIN THE COEFFICIENT OF VARIATION PERFORMANCE FOR THE ASSAY. NO ERRONEOUS RESULTS WERE GENERATED. NO ADVERSE EVENTS WERE REPORTED.
THE USER EXPERIENCED AN INTERMITTENT ISSUE WITH QUESTIONABLE TROPONIN T AND TROPONIN T STAT GENERATION 4 (TROPONIN T STAT) RESULTS THAT WAS ONGOING SINCE (B)(6). THE USER PROVIDED DATA FOR ONE PATIENT SAMPLE. THE SAMPLE WAS INITIALLY TESTED ON COBAS E411 SERIAL NUMBER (B)(4) AT A SISTER SITE. THE INITIAL TROPONIN T RESULT WAS 0.086 NG/ML WITH A DATA FLAG AND THE INITIAL TROPONIN T STAT RESULT WAS 0.021 NG/ML. THE REPEAT TROPONIN T RESULT WAS 0.038 NG/ML WITH A DATA FLAG AND THE REPEAT TROPONIN T STAT RESULT WAS 0.023 NG/ML. THE SAMPLE WAS THEN TESTED ON THE E411 AT THIS SITE. THE TROPONIN T RESULT WAS 0.016 NG/ML AND THE TROPONIN T STAT RESULT WAS 0.023 NG/ML. A RESULT OF 0.02 NG/ML WAS REPORTED FOR THIS SAMPLE AT THE SISTER SITE. THE RESULTS FROM THIS ANALYZER WERE NOT REPORTED OUTSIDE THE LABORATORY AS THE SAMPLE WAS ONLY TESTED AS VERIFICATION FOR THEIR SISTER HOSPITAL. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE TROPONIN T REAGENT LOT NUMBER WAS 15930001 AND THE TROPONIN T STAT REAGENT LOT NUMBER WAS 15989402. THE USER DECLINED A SERVICE VISIT AS SHE STATED THE SAMPLE WAS WITHIN THEIR IN-HOUSE PRECISION GUIDELINES AND SHE DID NOT BELIEVE THERE WAS AN ANALYZER ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS E411 DISK SYSTEM | IMMUNOCHEMISTRY ANALYZER | MMI | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 062 YR |