COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2011-00092
- Event Type
- Malfunction
- Date Received
- February 25, 2011
- Date of Event
- January 28, 2011
- Report Date
- January 28, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BD
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE COMPLAINT CASTOR FROM A PLASTIC BASE WAS RETURNED AND VISUALLY INSPECTED. IN ADDITION THE FOOT CLIP SECURITY PLATE WAS ALSO RETURNED. RESULTS: THE VISUAL INSPECTION REVEALED THAT THE THREADED BOLT STEAM OF THE COMPLAINT CASTOR HAD SNAPPED OFF AFTER THE LOCKING NUT. THE CORRESPONDING PIECE RECEIVED WAS 45 MM IN LENGTH AND PRESENTED WITH THE SAME STRESS MARKS AS THE OTHER PIECE. THE WHEELS WERE OBSERVED TO HAVE SOME FLAT SPOTS AND VARIOUS SCRATCHES ON IT. SOME DENTS WERE ALSO OBSERVED ON THE LEADING FRONT OF THE CASTOR POLYAMIDE BODY. THE VISUAL INSPECTION ALSO REVEALED A DENT ON THE RETURNED FOOT CLIP SECURITY PLATE, AND THAT IT WAS SLIGHTLY BENT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: BASED ON THE INSPECTION OF THE RETURNED COMPLAINT DEVICES, IT IS LIKELY THAT THE CASTOR HAD SHEARED OFF DUE TO IMPACT DAMAGE. THE DENTS OBSERVED ON THE CASTOR BODY AND THE BENT CLIP SECURITY PLATE IMPLIES THAT THE WHEEL HAS BEEN SUBJECTED TO SOME FORM OF IMPACT FORCE. THE MOBILE INFANT WARMER PROVIDES MOBILE WARMING WHEREVER NEEDED. IT CAN BE SUSCEPTIBLE TO IMPACT DAMAGE WHEN IT IS MOVED FROM ONE ROOM TO ANOTHER FOR VARIOUS REASONS, SUCH AS ACCIDENTAL BUMPS AND/OR COLLISIONS BETWEEN THE CASTOR WHEELS AND WALLS, UNEVEN SURFACES AND TRANSPORTING VIA ELEVATORS. THE LOT DATE INDICATES THAT THIS DEVICE IS SIX YEARS OLD ,AND THAT IT IS POSSIBLE FOR DYNAMIC STRESSES TO OCCUR AT THE BASE REGION OF INFANT WARMERS OVER A PERIOD OF TIME FROM A COMBINATION OF OVERLOADING AND TRANSPORT TO AND FROM DIFFERENT WARDS THE PRODUCT TECHNICAL MANUAL ADVISES THAT THE CASTORS SHOULD BE CHECKED TO ENSURE THAT THEY ROLL AND LOCK PROPERLY AS PART OF THE ANNUAL FUNCTIONAL CHECK. A REPLACEMENT METAL BASE HAS BEEN SUPPLIED TO THE CUSTOMER.
(B)(4). WE ARE CURRENTLY ENDEAVOURING TO OBTAIN MORE INFORMATION FROM THE CUSTOMER WITH REGARD TO THE COMPLAINT DEVICE. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE REAR RIGHT CASTOR OF AN IW934 COSYCOT INFANT WARMER "FELL OFF" WHILE STEERING THE DEVICE. THIS WAS FOUND PRIOR TO PATIENT USE.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE REAR RIGHT CASTOR OF AN (B)(4) COSYCOT INFANT WARMER "FELL OFF" WHILE STEERING THE DEVICE. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COSYCOT INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW934JEU | 050305 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |