SYNCHRON CX5 DELTA CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-00544
- Event Type
- Injury
- Date Received
- February 24, 2011
- Date of Event
- January 24, 2011
- Report Date
- January 24, 2011
- Manufacturer
- BECKMAN COULTER INC
- Product Code
- JJE
- PMA / PMN Number
- K881495
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE ARE IN PLASMA TUBE WHICH WERE SPUN FOR 10 MINUTES AND POURED INTO AN INSERT CUP. SAMPLE WAS SLIGHTLY ICTERIC. PER CUSTOMER, QC PRIOR TO AND AFTER THE EVENT WAS WITHIN THE LAB'S ESTABLISHED RANGES. THIS IS A SAMPLE SPECIFIC EVENT; THEREFORE, SERVICE CALL WAS NOT GENERATED OR INITIATED. A CLEAR ROOT CAUSE IS UNKNOWN FOR THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO LOW GLUCOSE (GLU) CARTRIDGE RESULT GENERATED BY SYNCHRON CX5 DELTA CLINICAL SYSTEM. THE ERRONEOUS RESULT WAS NOT REPORTED OUT OF THE LABORATORY; HENCE NO PATIENT TREATMENT WAS IMPACTED. THE ORIGINAL SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND AN ALTERNATE INSTRUMENT AND HIGHER RESULTS WERE OBTAINED. IN ADDITION, THE SAMPLE WAS RUN ON SYNCHRON CX9 PRO CLINICAL SYSTEM AND GLUCOSE MODULAR RESULT WAS 45 MG/DL. THE SAMPLE WAS REPEATED AND A RESULT OF 47 MG/DL WAS OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON CX5 DELTA CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC | CX 5 DELTA | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |