FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2001283
·
Received February 24, 2011
Report
- Report Number
- 2050012-2011-00559
- Event Type
- Malfunction
- Date Received
- February 24, 2011
- Date of Event
- January 26, 2011
- Report Date
- January 26, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BCI FIELD SERVICE ENGINEER (FSE) CONTACTED THE CUSTOMER ON (B)(4) 2011. DURING TROUBLESHOOTING IT WAS DISCOVERED THAN THE VACUUM LINE FROM THE ISE DRAIN WAS LOOSE. THE CUSTOMER RECONNECTED THE LINE AND INSPECTED FOR LEAKS. NO LEAK WAS OBSERVED. SERVICE WAS NOT DISPATCHED FOR THIS EVENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO A LEAKAGE ON UNICEL DXC 600 SYNCHRON CHEMISTRY ANALYZER. THE CUSTOMER STATED THE INSTRUMENT WAS LEAKING ONTO MODULAR CHEMISTRY SIDE DRIP TRAY AFTER CHANGING CO2 ALKALINE BUFFER. NO OPERATOR EXPOSURE WAS NOTED. BCI CUSTOMER TECHNICAL SUPPORT RECOMMENDED REVIEW OF WASH CONCENTRATE MSDS. THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |