ADULT BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00094
- Event Type
- Malfunction
- Date Received
- February 24, 2011
- Date of Event
- January 27, 2011
- Report Date
- January 31, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K033710
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). (B)(6). METHOD: THE RETURNED BREATHING CIRCUIT WAS PRESSURE TESTED AND SUBMERGED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: A LEAK WAS FOUND IN THE INSPIRATORY AND EXPIRATORY WATER TRAPS. THE WATER TRAP OF THE BREATHING CIRCUIT CONSISTS OF TWO PARTS WHERE THE LOWER PART CAN BE REMOVED DURING USE FOR DRAINING WATER. THE WATER TRAP LEAKS WERE LOCATED AT THE SEAL BETWEEN THE WATER TRAP BOWL AND THE WATER TRAP LID. A LOT CHECK REVEALED NO OTHER SIMILAR COMPLAINTS FOR LOT NUMBER 100629. CONCLUSION: ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED POST PRODUCTION DURING TRANSPORT, STORAGE OR SETUP. OUR MONITORING AND TRENDING OF EVENTS INVOLVING WATER TRAP LEAKS IN ADULT BREATHING CIRCUITS HAS A RATE OF OCCURRENCE OF (B)(4).
(B)(4). THE COMPLAINT DEVICE HAS NOT YET BEEN RECEIVED BY FISHER & PAYKEL HEALTHCARE ((B)(4)) FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THERE WAS AN AIR LEAK FROM THE WATER TRAP OF AN RT134 ADULT BREATHING CIRCUIT.THE LEAK WAS OBSERVED DURING SETUP OF THE DEVICE.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THERE WAS AN AIR LEAK FROM THE WATER TRAP OF AN RT134 ADULT BREATHING CIRCUIT. THE LEAK WAS OBSERVED DURING SETUP OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT134 | 100629 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |