FDA Adverse Event Injury Summary report: N

INGENIA 1.5T

MDR report key: 20012445 · Received August 21, 2024

Report

Report Number
3003768277-2024-04554
Event Type
Injury
Date Received
August 21, 2024
Date of Event
June 3, 2024
Report Date
August 20, 2024
Manufacturer
PHILIPS ELECTRONICS NEDERLAND B.V.
Product Code
LNH
UDI-DI
00884838055322
PMA / PMN Number
K163116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION: THERE IS NO INDICATION OF A MALFUNCTION OF THE MR SYSTEM OR COILS USED THAT COULD HAVE CONTRIBUTED TO THE INCIDENT. THE INJURY, A LARGE 2ND DEGREE BURN TO THE LEFT SIDE OF THE CHEST CAN BE EXPLAINED BY INSUFFICIENT DISTANCE BETWEEN BODY PARTS, A SO-CALLED SKIN-TO-SKIN BURN BETWEEN HER LEFT CHEST AND LEFT UPPER ARM. CONTRIBUTING FACTORS: THE THERMOREGULATION OF OBESE PATIENTS IS KNOWN TO BE IMPAIRED. THE RISK OF RF ENERGY-RELATED INJURIES IS HIGHER IN PATIENTS WITH IMPAIRED THERMOREGULATION. THE THERMOREGULATION OF PATIENTS WHO ARE HYPERTENSIVE IS OFTEN IMPAIRED. THE RISK OF RF ENERGY-RELATED INJURIES IS HIGHER IN PATIENTS WITH IMPAIRED THERMOREGULATION. THE TOTAL ADMINISTERED SPECIFIC ENERGY DOSE OF 4.3 KJ/KG IS VERY HIGH FOR A PATIENT WITH IMPAIRED THERMOREGULATION. THE INSTRUCTIONS FOR USE ADVISES TO AVOID SED >3.0 KJ/KG, PREFERABLY KEEP SED <2.0 KJ/KG FOR THOSE PATIENTS. THE VENTILATION WAS ON LEVEL 3. LEVEL 5 IS RECOMMENDED, IT SUPPORTS THE COOL DOWN MECHANISM.

Description of Event or Problem · 0

PHILIPS RECEIVED A REPORT THROUGH THE MEDWATCH SYSTEM THAT A PATIENT SUFFERED A LARGE 2ND DEGREE BURN TO THE LEFT SIDE OF THE CHEST DURING A NECK AND SPINE EXAMINATION. THE PATIENT REQUIRED TREATMENT AS A RESULT OF THIS INJURY SUSTAINED. THE CUSTOMER DID NOT REPORT THIS INCIDENT THROUGH OUR REGULAR SERVICE CHANNELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1153513 INGENIA 1.5T SYSTEM, NUCLEAR MAGNETIC IMAGING LNH PHILIPS ELECTRONICS NEDERLAND B.V. INGENIA 1.5T R5 BASED 00884838055322

Patients

Seq Age Sex Outcome Treatment
1 52 YR Female Required Intervention| O