TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-00336
- Event Type
- Injury
- Date Received
- February 24, 2011
- Date of Event
- January 25, 2011
- Report Date
- January 26, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- NIQ
- PMA / PMN Number
- P060008
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS A COMBINATION PRODUCT. (B)(4).
DEVICE EVALUATED BY MANUFACTURER - RETURNED PRODUCT CONSISTED OF A TAXUS LIBERTE (MR) STENT DELIVERY SYSTEM (SDS) WITH NO STENT OR SNARE. THERE WERE STENT STRUT IMPRESSIONS PRESENT ON THE SURFACE OF THE BALLOON BETWEEN THE MARKER BANDS, INDICATING THE STENT WAS APPROPRIATELY POSITIONED AND SECURED TO THE BALLOON IN MANUFACTURING. THE BALLOON WAS IN A TIGHTLY FOLDED CONDITION, AS-RECEIVED, AND MICROSCOPIC EXAMINATION OF THE BALLOON PRESENTED NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES. IT WAS DETERMINED THE TIP WAS DAMAGED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
SAME CASE AS MFR # 2134265-2011-00413 AND 2134265-2011-00434. SAME CASE AS USER FACILITY MEDWATCH # (B)(4). IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, A STENT DISLODGEMENT OCCURRED. THE 90% LESION BEING TREATED WAS LOCATED IN THE SEVERELY DISEASED SAPHENOUS VEIN GRAFT (SVG) TO THE OBTUSE MARGINAL (OM). WITHIN THE VESSEL THERE WERE 2 UNIDENTIFIED PREVIOUSLY IMPLANTED PROXIMAL STENTS, FOLLOWED BY THE LESION BEING TREATED, AND A DISTAL UNIDENTIFIED STENT. A FILTERWIRE AND IQ WIRE WERE PLACED, USING A BUDDY WIRE TECHNIQUE. DIFFICULTY PLACING BOTH WIRES TOOK AN "EXCEEDINGLY LONG TIME" AS THE WIRES WERE GETTING CAUGHT ON AREAS THROUGHOUT THE VEIN GRAFT. HOWEVER, AFTER PLACEMENT OF THE FILTERWIRE (PRIOR TO USE OF THE STENT), THERE WAS 'ESSENTIALLY' NO FLOW IN THE VEIN GRAFT. THE LESION WAS PREDILATED WITH A 3.5X20MM APEX BALLOON, RESTORING FLOW. A SIGNIFICANT DAMPENING OF THE PRESSURE WAVE FORM WAS IDENTIFIED. THE PHYSICIAN ATTEMPTED TO PLACE A 3.5X28MM TAXUS LIBERTE IN THE MID PORTION, BUT WAS UNABLE TO CROSS THE MID PORTION. THEREFORE THE PHYSICIAN DEPLOYED THE STENT FROM THE OSTIUM TO THE PROXIMAL PORTION, INFLATED 4 TIMES TO 12-18ATMS FOR 18-29 SECONDS. AFTER PLACEMENT OF THIS STENT THE DAMPENING OF THE PRESSURE WAVE FORM WAS NOT SEEN. THE PHYSICIAN ATTEMPTED TO PLACE ANOTHER 3.5X38MM TAXUS LIBERTE STENT, BUT WAS UNABLE TO CROSS THE MID PORTION OF THE LESION. THE FILTERWIRE WAS THEN REMOVED AS IT HAD LOST POSITIONING DURING ONE OF THE STENT PASSES. THE FILTERWIRE BECAME WRAPPED UP WITH THE FIRST IQ WIRE AND THE ENTIRE SYSTEM HAD TO BE REMOVED AND A NEW IQ WIRE WAS PLACED. BALLOON DILATATION WAS PERFORMED WITH THE SAME 3.5X20MM APEX BALLOON; TOTAL OF 4 INFLATIONS WERE MADE FROM 12-14ATMS FOR 15-22 SECONDS. THE PHYSICIAN ATTEMPTED TO PLACE THE SAME 3.5X38MM TAXUS LIBERTE STENT. HOWEVER, THE STENT WOULD NOT PASS THROUGH THE PROXIMAL STENT. THE PROXIMAL STENT WAS DILATED AGAIN WITH A 4.0X30MM QUANTUM APEX, 3 TIMES TO 20ATMS 18-33 SECONDS. THE PHYSICIAN ATTEMPTED TO PLACE THE SAME 3.5X38MM TAXUS LIBERTE STENT AND WAS ABLE TO REACH THE MID PORTION BUT COULD NOT ADVANCE ANY FURTHER. UPON WITHDRAWAL THE STENT CAUGHT ON THE PREVIOUSLY PLACED PROXIMAL STENT AND DISLODGED FROM THE STENT DELIVERY SYSTEM (SDS). FLOW REMAINED ADEQUATE EVEN THOUGH THE DISLODGED STENT WAS SOMEWHAT CRUSHED TOWARD THE PROXIMAL PORTION. THE PHYSICIAN ATTEMPTED TO INFLATE THE DISLODGED STENT WITH A 4.0X20MM QUANTUM APEX, BUT WAS UNABLE TO CROSS THROUGH THE STENT. A 3.0X20MM APEX WAS ABLE TO INFLATE THE STENT, 3 INFLATIONS WERE MADE TO 14ATMS FOR 18-21 SECONDS. THEN THE PREVIOUS 4.0X20MM QUANTUM APEX WAS USED TO FURTHER INFLATE BOTH OF THE TAXUS LIBERTE STENTS, 5 INFLATIONS WERE MADE TO 20ATMS FOR 13-18 SECONDS. THERE WAS EXCELLENT FLOW IN THE VESSEL. HOWEVER, THERE WAS HAZINESS AT THE MAIN LESION. A 3.0X18MM PROMUS STENT WAS IMPLANTED IN THE MID PORTION, INFLATED TO 16ATMS FOR 30 SECONDS. A 3.0X23MM PROMUS STENT WAS ALSO IMPLANTED IN THE PROXIMAL TO MID PORTION, INFLATED TO 16ATMS FOR 22 SECONDS. THE STENTS WERE POST DILATED WITH A 4.0X30MM QUANTUM BALLOON, INFLATED 7TIMES TO 20ATMS FOR 15-27 SECONDS. LESS THAN 10% RESIDUAL STENOSIS REMAINS THROUGHOUT THE STENTED AREA. TIMI FLOW IS 3. MEDICATIONS GIVEN INCLUDED: HEPARIN AND ANGIOMAX. NO FURTHER PATIENT COMPLICATIONS REPORTED AND THE PATIENT IS FINE. PER THE PHYSICIAN, "IT WAS A VERY EXTRAORDINARILY COMPLEX PROCEDURE, BUT ULTIMATELY SUCCESSFUL."
SAME CASE AS MFR # 2134265-2011-00413 AND 2134265-2011-00434. SAME CASE AS USER FACILITY MEDWATCH # (B)(4). IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, A STENT DISLODGEMENT OCCURRED. THE 90% LESION BEING TREATED WAS LOCATED IN THE SEVERELY DISEASED SAPHENOUS VEIN GRAFT (SVG) TO THE OBTUSE MARGINAL (OM). WITHIN THE VESSEL, THERE WERE 2 UNIDENTIFIED PREVIOUSLY IMPLANTED PROXIMAL STENTS, FOLLOWED BY THE LESION BEING TREATED, AND A DISTAL UNIDENTIFIED STENT. A FILTERWIRE AND IQ WIRE WERE PLACED, USING A BUDDY WIRE TECHNIQUE. DIFFICULTY PLACING BOTH WIRES TOOK AN "EXCEEDINGLY LONG TIME" AS THE WIRES WERE GETTING CAUGHT ON AREAS THROUGHOUT THE VEIN GRAFT. HOWEVER, AFTER PLACEMENT OF THE FILTERWIRE (PRIOR TO USE OF THE STENT), THERE WAS 'ESSENTIALLY' NO FLOW IN THE VEIN GRAFT. THE LESION WAS PREDILATED WITH A 3.5X20MM APEX BALLOON, RESTORING FLOW. A SIGNIFICANT DAMPENING OF THE PRESSURE WAVE FORM WAS IDENTIFIED. THE PHYSICIAN ATTEMPTED TO PLACE A 3.5X28MM TAXUS LIBERTE IN THE MID PORTION, BUT WAS UNABLE TO CROSS THE MID PORTION. THEREFORE, THE PHYSICIAN DEPLOYED THE STENT FROM THE OSTIUM TO THE PROXIMAL PORTION, INFLATED 4 TIMES TO 12-18ATMS FOR 18-29 SECONDS. AFTER PLACEMENT OF THIS STENT, THE DAMPENING OF THE PRESSURE WAVE FORM WAS NOT SEEN. THE PHYSICIAN ATTEMPTED TO PLACE ANOTHER 3.5X38MM TAXUS LIBERTE STENT, BUT WAS UNABLE TO CROSS THE MID PORTION OF THE LESION. THE FILTERWIRE WAS THEN REMOVED AS IT HAD LOST POSITIONING DURING ONE OF THE STENT PASSES. THE FILTERWIRE BECAME WRAPPED UP WITH THE FIRST IQ WIRE AND THE ENTIRE SYSTEM HAD TO BE REMOVED AND A NEW IQ WIRE WAS PLACED. BALLOON DILATATION WAS PERFORMED WITH THE SAME 3.5X20MM APEX BALLOON; TOTAL OF 4 INFLATIONS WERE MADE FROM 12-14ATMS FOR 15-22 SECONDS. THE PHYSICIAN ATTEMPTED TO PLACE THE SAME 3.5X38MM TAXUS LIBERTE STENT. HOWEVER, THE STENT WOULD NOT PASS THROUGH THE PROXIMAL STENT. THE PROXIMAL STENT WAS DILATED AGAIN WITH A 4.0X30MM QUANTUM APEX, 3 TIMES TO 20ATMS 18-33 SECONDS. THE PHYSICIAN ATTEMPTED TO PLACE THE SAME 3.5X38MM TAXUS LIBERTE STENT AND WAS ABLE TO REACH THE MID PORTION BUT COULD NOT ADVANCE ANY FURTHER. UPON WITHDRAWAL THE STENT CAUGHT ON THE PREVIOUSLY PLACED PROXIMAL STENT AND DISLODGED FROM THE STENT DELIVERY SYSTEM (SDS). FLOW REMAINED ADEQUATE EVEN THOUGH THE DISLODGED STENT WAS SOMEWHAT CRUSHED TOWARD THE PROXIMAL PORTION. THE PHYSICIAN ATTEMPTED TO INFLATE THE DISLODGED STENT WITH A 4.0X20MM QUANTUM APEX, BUT WAS UNABLE TO CROSS THROUGH THE STENT. A 3.0X20MM APEX WAS ABLE TO INFLATE THE STENT, 3 INFLATIONS WERE MADE TO 14ATMS FOR 18-21 SECONDS. THEN THE PREVIOUS 4.0X20MM QUANTUM APEX WAS USED TO FURTHER INFLATE BOTH OF THE TAXUS LIBERTE STENTS, 5 INFLATIONS WERE MADE TO 20ATMS FOR 13-18 SECONDS. THERE WAS EXCELLENT FLOW IN THE VESSEL. HOWEVER, THERE WAS HAZINESS AT THE MAIN LESION. A 3.0X18MM PROMUS STENT WAS IMPLANTED IN THE MID PORTION, INFLATED TO 16ATMS FOR 30 SECONDS. A 3.0X23MM PROMUS STENT WAS ALSO IMPLANTED IN THE PROXIMAL TO MID PORTION, INFLATED TO 16ATMS FOR 22 SECONDS. THE STENTS WERE POST DILATED WITH A 4.0X30MM QUANTUM BALLOON, INFLATED 7 TIMES TO 20ATMS FOR 15-27 SECONDS. LESS THAN 10% RESIDUAL STENOSIS REMAINS THROUGHOUT THE STENTED AREA. TIMI FLOW IS 3. MEDICATIONS GIVEN INCLUDED: HEPARIN AND ANGIOMAX. NO FURTHER PATIENT COMPLICATIONS REPORTED AND THE PATIENT IS FINE. PER THE PHYSICIAN, "IT WAS A VERY EXTRAORDINARILY COMPLEX PROCEDURE, BUT ULTIMATELY SUCCESSFUL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - MAPLE GROVE | H7493893638350 | 13819778 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Required Intervention | 3.5X38MM TAXUS LIBERTE LONG STENT| 3.5X20MM APEX BALLOON| FILTERWIRE 190| IQ GUIDE WIRE |