FDA Adverse Event Other Summary report: N

IBV VALVE SYSTEM

MDR report key: 2001117 · Received February 17, 2011

Report

Report Number
3004450998-2011-00001
Event Type
Other
Date Received
February 17, 2011
Date of Event
February 9, 2011
Report Date
February 17, 2011
Manufacturer
SPIRATION, INC.
Product Code
OAZ
PMA / PMN Number
H060002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SPIRATION REVIEWED DEVICE LABELING (INSTRUCTIONS FOR USE AND PATIENT BROCHURE) AND CONFIRMED THAT: THE ADVERSE EVENT (PNEUMOTHORAX) WAS LISTED AS POTENTIAL/ANTICIPATED, AND THE USE OF THE IBV VALVES WAS OFF-LABEL. METHOD: REVIEWED THE (B)(4) REPORT SENT BY THE PHYSICIAN TO HIS IRB ON (B)(6) 2011. BASED ON THE REVIEW OF THE (B)(4) REPORT, THE IBV VALVE IS NOT SUSPECTED TO HAVE FAILED OR MALFUNCTIONED. THIS REPORT IS BEING SUBMITTED AS AN MDR WITH AN ABUNDANCE OF CAUTION. IF ADDITIONAL SIGNIFICANT INFO IS PROVIDED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

PT (B)(6) UNDERWENT A BRONCHOSCOPY PROCEDURE PERFORMED BY DR. (B)(6) ON (B)(6) 2011 FOR IBV VALVE PLACEMENT (OFF-LABEL USE) TO TREAT THE GIANT BULLAE IN THE LEFT UPPER LOBE. A POST-PROCEDURE CHEST X-RAY ON (B)(6) 2011 COULD NOT RULE OUT A PNEUMOTHORAX; HOWEVER, THE PT WAS HEMODYNAMICALLY STABLE. ON (B)(6) 2011, A REPEAT CHEST X-RAY WAS PERFORMED WHICH ALSO COULD NOT CONFIRM OR RULE OUT PNEUMOTHORAX. A CT SCAN WAS THEN REQUESTED AND A PNEUMOTHORAX WAS VISIBLE WITH LEFT LOWER LOBE COLLAPSE AND MEDIASTINAL SHIFT TO THE RIGHT. THE PT BECAME TACHYCARDIC, HYPOTENSIVE, AND WAS IN RESPIRATORY DISTRESS. A CHEST TUBE WAS PLACED AND CONNECTED TO WALL SUCTION WITH CLINICAL AND RADIOGRAPHICAL IMPROVEMENT. THE (B)(4) REPORT TO THE IRB DATED (B)(6) 2011 STATED THAT: THE EVENT WAS PROBABLY RELATED TO THE DEVICE BASED ON THE TIMELINE OF THE EVENT WITHIN 24 HOURS OF THE PROCEDURE, THE EVENT DID NOT INCREASE THE LIKELY RISK OR DECREASE THE LIKELY BENEFIT TO PTS, AND THIS RISK IS EXPECTED AS IT IS LISTED IN THE PT BROCHURE GIVEN TO PT DURING THE CONSENT PROCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IBV VALVE SYSTEM ONE-WAY AIR LEAK VALVE OAZ SPIRATION, INC. W04036-01

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L