FDA Adverse Event Injury Summary report: N

100CC SILICONE BULB EVACUATOR

MDR report key: 2000977 · Received February 18, 2011

Report

Report Number
1018233-2011-00019
Event Type
Injury
Date Received
February 18, 2011
Date of Event
January 14, 2011
Report Date
January 21, 2011
Manufacturer
C.R. BARD, INC.
Product Code
GBX
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD WAS REVIEWED FINDING NOTHING THAT COULD CAUSE OR CONTRIBUTE TO THE REPORTED FAILURE. THE SAMPLE WAS RETURNED FOR EVALUATION FINDING THE FLAT DRAIN TUBE BROKEN AT THE FIRST EYE AT THE BEGINNING OF THE TRANSITION. THE SAMPLE WAS EXAMINED UNDER MAGNIFICATION FINDING NO EVIDENCE OF ANY PUNCTURES. THE BROKEN SURFACE HAD JAGGED EDGES WHICH IS INDICATIVE OF EXCESSIVE FORCE HAVING BEEN APPLIED TO THE DRAIN BEYOND ITS TENSILE CAPABILITIES. THERE IS NOTHING IN THE MANUFACTURING PROCESS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED FAILURE. THE INSTRUCTIONS FOR USE STATES THE FOLLOWING PRECAUTIONS "ADDITIONAL PERFORATIONS SHOULD NOT BE MADE IN THE DRAIN. AVOID SUTURING THROUGH DRAIN. DRAIN SHOULD LIE FLAT AND IN LINE WITH THE SKIN EXIT AREAS. PARTICULAR CARE SHOULD BE TAKEN TO AVOID ANY OBSTACLES TO THE DRAIN EXIT PATH. DRAIN SHOULD BE CHECKED FOR FREE MOTION DURING CLOSURE TO MINIMIZE THE POSSIBILITY OF BREAKAGE. DRAIN REMOVAL SHOULD BE DONE GENTLY BY HAND. DRAIN SHOULD NOT BE HANDLED WITH POINTED, TOOTHED OR SHARP INSTRUMENTS WHICH COULD CAUSE CUTS OR NICKS AND LEAD TO SUBSEQUENT STRUCTURAL FAILURE OF THE DRAIN." AN ADDITIONAL WARNING STATES: "DRAIN BREAKAGE MAY REQUIRE SURGICAL REMOVAL." (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN ATTEMPTING TO REMOVE THE DRAIN FROM THE LEFT KIDNEY, THE DRAIN BROKE. THERE WAS MINIMAL RESISTANCE REPORTED WHILE REMOVING THE DRAIN. THE PATIENT WAS TAKEN TO SURGERY FOR REMOVAL OF THE RETAINED PIECE. THE PORTION OF THE DRAIN REMOVED WAS MATCHED TO THE REMAINING PORTION OF THE DRAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 100CC SILICONE BULB EVACUATOR GBX C.R. BARD, INC. NA NGUG2323

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention