MDR report key: 2000975 · Received February 18, 2011

Report

Report Number
3006630150-2011-00250
Event Type
Injury
Date Received
February 18, 2011
Date of Event
December 15, 2010
Report Date
December 15, 2010
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P030017
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE IPG PASSED VISUAL, PHOTOGRAPHIC IMAGING, PERFORMANCE AND ELECTRICAL TESTS PERFORMED. THE DEVICE OUTPUT WAS MONITORED FOR EXCESSIVE LEAKAGE CURRENT OR SPURIOUS SIGNALS IN A SALINE SOLUTION, WHILE PROGRAMMED TO THE PATIENT'S LATEST SETTING. NO DISCREPANCIES WERE IDENTIFIED. THE DEVICE EXHIBITED NORMAL DEVICE CHARACTERISTICS. THE LEADS PASSED VISUAL AND PHOTOGRAPHIC IMAGING TESTS PERFORMED. THE LEADS EXHIBITED NORMAL DEVICE CHARACTERISTICS.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT A PATIENT WAS FEELING 'JERKING SENSATION' IN HER CERVICAL IMPLANT. DURING A LEAD REVISION SURGERY THE PATIENT REPORTED DISCOMFORT AT THE BATTERY SITE. THE PHYSICIAN EXPLANTED THE WHOLE PRECISION SYSTEM AS THE IMPEDANCES WERE SHOWING REALLY LOW. THE PATIENT IS REPORTEDLY DOING WELL AFTER THE PROCEDURE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT A PATIENT WAS FEELING 'JERKING SENSATION' IN HER CERVICAL IMPLANT. DURING A LEAD REVISION SURGERY THE PATIENT REPORTED DISCOMFORT AT THE BATTERY SITE. THE PHYSICIAN EXPLANTED THE WHOLE PRECISION SYSTEM AS THE IMPEDANCES WERE SHOWING REALLY LOW. THE PATIENT IS REPORTEDLY DOING WELL AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2218-70 NA

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention ST LINEAR LEAD, 70CM WITH PRE-LOADED 0.014" STYLET| MODEL# SC-2218-70, SERIAL# (B)(4)| IPG KIT (WITHOUT PULL-THROUGH TUNNELER):| MODEL# SC-1110-02, SERIAL# (B)(4)