Report Number
2953200-2011-00417
Event Type
Injury
Date Received
February 16, 2011
Date of Event
January 19, 2011
Report Date
January 19, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: (ENDOLEAK). EVALUATION, RESULTS/CONCLUSION: (MODERATELY ANGULATED AND THROMBOSIS AORTIC NECK, AND HAD AN AREA OF 3-4MM X 2MM CALCIFICATION IN THE POSTERIOR AORTIC NECK).

Description of Event or Problem · 1

A TALENT STENT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.5 CM THORACIC AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 23.5 MM IN DIAMETER AT THE RENAL ARTERIES AND 25 MM IN DIAMETER ABOVE THE AAA, MODERATELY ANGULATED, HAD THROMBUS, AND HAD AN AREA OF 3-4MM X 2MM CALCIFICATION IN THE POSTERIOR NECK. THE ILIAC ARTERIES WERE CALCIFIED. IT WAS REPORTED THAT AFTER THE INITIAL IMPLANTATION OF THE GRAFTS AND MODELING WITH A RELIANT BALLOON, A POST ANGIOGRAM SHOWED A SIGNIFICANT PROXIMAL TYPE I ENDOLEAK; HOWEVER, REPEAT BALLOONING DID NOT IMPROVE THE LEAK. THE PHYSICIAN EXTENDED PROXIMALLY BY 3-4MM WITH A CUFF (REF MFR 2953200-2011-00418), WHICH GREATLY IMPROVED THE LEAK, BUT STILL A PERSISTENT PROXIMAL TYPE I ENDOLEAK REMAINED. NO FURTHER INTERVENTION WAS PERFORMED, AND THE PT WILL BE MONITORED WITH A 30 DAY CT FOLLOW-UP. WILL MONITOR IN 30 DAYS. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA NA V00797508

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention