TALENT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-00417
- Event Type
- Injury
- Date Received
- February 16, 2011
- Date of Event
- January 19, 2011
- Report Date
- January 19, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (ENDOLEAK). EVALUATION, RESULTS/CONCLUSION: (MODERATELY ANGULATED AND THROMBOSIS AORTIC NECK, AND HAD AN AREA OF 3-4MM X 2MM CALCIFICATION IN THE POSTERIOR AORTIC NECK).
A TALENT STENT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.5 CM THORACIC AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 23.5 MM IN DIAMETER AT THE RENAL ARTERIES AND 25 MM IN DIAMETER ABOVE THE AAA, MODERATELY ANGULATED, HAD THROMBUS, AND HAD AN AREA OF 3-4MM X 2MM CALCIFICATION IN THE POSTERIOR NECK. THE ILIAC ARTERIES WERE CALCIFIED. IT WAS REPORTED THAT AFTER THE INITIAL IMPLANTATION OF THE GRAFTS AND MODELING WITH A RELIANT BALLOON, A POST ANGIOGRAM SHOWED A SIGNIFICANT PROXIMAL TYPE I ENDOLEAK; HOWEVER, REPEAT BALLOONING DID NOT IMPROVE THE LEAK. THE PHYSICIAN EXTENDED PROXIMALLY BY 3-4MM WITH A CUFF (REF MFR 2953200-2011-00418), WHICH GREATLY IMPROVED THE LEAK, BUT STILL A PERSISTENT PROXIMAL TYPE I ENDOLEAK REMAINED. NO FURTHER INTERVENTION WAS PERFORMED, AND THE PT WILL BE MONITORED WITH A 30 DAY CT FOLLOW-UP. WILL MONITOR IN 30 DAYS. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA | NA | V00797508 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |