ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2000579 · Received February 1, 2011

Report

Report Number
2531779-2011-00576
Event Type
Malfunction
Date Received
February 1, 2011
Report Date
January 3, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED. THE "UP/DOWN/OK/CONTRAST" KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION. THE "UP/DOWN/OK/CONTRAST" KEY CONTACTS WERE OBSERVED TO BE INVERTED. EVIDENCE OF KEYPAD ADHESIVE WAS FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PT CLAIMED THAT THE PUMP WAS INTERMITTENTLY NOT RESPONDING TO BUTTON PRESSES. THE PT DENIED ANY TRAUMA TO THE KEYPAD. SHE ALSO DENIED THAT THE DEVICE WAS EXPOSED TO WATER. THE PT CLAIMED THAT THE KEYPAD WAS INTACT AND WAS NOT PEELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA